Productomschrijving
TH-ICH-300 Specific Purpose Chamber. -
Chamber volume (L ):300 . -
Controller: PID control (LCD Type). -
Temperature Range without humidity(°C): -5 to 85 without Light / 0 to 85 with Light. -
Temperature Range with humidity(°C): +15 to 85 with Light . -
Temperature Fluctuation (±°C) at 40°C / 60%RH: 0.3. -
Temperature Variation (±°C) at 40°C / 60%RH: 0.5. -
Time Heating (min.): 25 from -20°C to 85°C. -
Time Cooling (min.): 35 from 20 °C to -5°C. -
Humidity Range (%RH): 25 to 95. -
Humidity Fluctuation (±%RH)at 60%RH / 40°C: 3. -
Humidity Variation (±%RH)at 60%RH / 40°C: 5.-
Light source (ICH Q1B option 2) Cool white: 6,000 lux. -
Light source (ICH Q1B option 2) Ultraviolet-A: 4.5 W/m2. -
Light source (ICH Q1B option 2) Light uniformity: VIS. : ±10%UVA : ±15%. -
Typical time required to reach ICH Recommendation of 1.2 million lux-hr: 200Hr. -
Typical time required to reach ICHrecommendation of 200 W-hr/m2: 50Hr. -
Refrigeration System: Air - cooled. -
Refrigeration Refrigerant: R- 404A. -
Water tank Capacity: 15L -
Water tank quality: pH6.2 to 7.2 / Electrical conductivity 20micro. sec. /cm to below. -
No. of wire shelves (standard / max.): 2 / 8. -
Distance of between shelves (mm): 60. -
Max. load per shelf (kg): 25. -
Permitted total load (kg): 75. -
Dimension (W×D×H) Interior (mm): 750×650×650. -
Dimension (W×D×H) Exterior (mm): 1130×985×1520. -
Net weight (kg): 280±10 . -
Cable port: Basically one hole provided / Ø50mm (2”) - Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”). -
Interface port: Basically RS-232C [RS-485-optional]. -
Electrical requirements AC230V, 1ph, 50Hz: 15.4 A.